REAL

Comparison of Four PCR and Two Point of Care Assays Used in the Laboratory Detection of SARS-CoV-2

Kenyeres, Bence and Ánosi, Noel and Bányai, Krisztián and Mátyus, Mária and Orosz, László and Kiss, Andrea and Kele, Beatrix and Burián, Katalin and Lengyel, György (2021) Comparison of Four PCR and Two Point of Care Assays Used in the Laboratory Detection of SARS-CoV-2. JOURNAL OF VIROLOGICAL METHODS : COVERING METHODOLOGY AND BASIC SCIENCE, 293. pp. 1-5. ISSN 0166-0934 (print); 1879-0984 (online)

[img] Text
Kenyeres_JVirMet_2021.pdf
Restricted to Repository staff only

Download (1MB)

Abstract

Seeing the global emergence and the lack of a definitive cure for COVID-19, it is essential to find the most sensitive and specific detection method to identify infected patients in a timely manner. Our paper aims to compare the clinical sensitivity of different commercial RT-qPCR (Genesig, 1copy, DNA-Technology and Charité primer-probe sets), isothermal PCR (Ustar Isothermal Amplification-Real Time Fluorescent Assay) and immunochromatographic antigen detection (BIOCREDIT COVID-19 Ag) assays developed to use in laboratory diagnosis of COVID-19. A total of 119 nasopharyngeal swab specimens were collected from symptomatic patients. A subset of samples, positive with two RT-qPCR assays were then tested with isothermal PCR and rapid antigen tests. Of the 119 specimens, 65 were positive by at least two PCR assays. All PCR assays showed substantial or perfect match, although some variations in the clinical performance was observed. Of the 37 and 32 remnant nasopharyngeal samples positive by RT-qPCR, respectively, three were positive by the BIOCREDIT COVID-19 Ag and 14 were detected by the isothermal amplification assay. In conclusion, in the clinical settings we recorded that each of the RT-qPCR assays was superior to other test formats, in particular, the routine use of the DNA-technology assay is recommended. Although alternative recommendations exist, we belive that the use of isothermal amplification assays and antigen rapid tests for COVID-19 diagnosis can only serve as adjuncts while awaiting the PCR result because of their high-false negative rate.

Item Type: Article
Subjects: Q Science / természettudomány > QR Microbiology / mikrobiológia > QR355 Virology / víruskutatás
Depositing User: dr. Enikő Fehér
Date Deposited: 17 Mar 2022 19:50
Last Modified: 17 Mar 2022 19:50
URI: http://real.mtak.hu/id/eprint/138923

Actions (login required)

Edit Item Edit Item