Balázs, Tibor and Merkely, Béla and Bognár, Eszter and Zima, Endre (2013) Interaction of Intraluminal Tissue and Coronary Sinus Lead Stabilized with Stent Placement. Journal of Cardiovascular Electrophysiology, 24 (4). pp. 468-470. ISSN 1045-3873, ESSN: 1540-8167
|
Text
JCE_cikk_10.1111-j.1540-8167.2012.02447.x másolata.pdf Download (231kB) | Preview |
Abstract
Intraluminal Interaction of Tissue and Stented CS Lead. Introduction: The aim of our investigation was to examine the intraluminal interaction of the vascular tissue and the implanted coronary sinus lead stabilized with stent on two human hearts removed before transplantation. Methods and results: The coronary sinus lumen was sectioned under operational microscope and opened carefully. The leads and stents were found separately positioned beside each other completely covered by an intact intimal tissue layer. No sign of occluding proliferative tissue was observed. Conclusion: Stent fixation technique and extraction of the CS lead in our cases did not have any particular damaging effect on the vascular system.
Item Type: | Article |
---|---|
Subjects: | T Technology / alkalmazott, műszaki tudományok > TA Engineering (General). Civil engineering (General) / általános mérnöki tudományok T Technology / alkalmazott, műszaki tudományok > TJ Mechanical engineering and machinery / gépészmérnöki tudományok |
Depositing User: | Dr. Eszter Bognár |
Date Deposited: | 25 Sep 2014 18:49 |
Last Modified: | 03 Apr 2023 08:19 |
URI: | http://real.mtak.hu/id/eprint/16849 |
Actions (login required)
Edit Item |