Véró, Mihály (1971) Negative capacitance amplifier for microelectrode investigations. A Magyar Tudományos Akadémia Tihanyi Biológiai Kutatóintézetének évkönyve, 38. pp. 107-115. ISSN 0365-3005
|
Text
109_cut_MTA_TihanyiBiologiaiKutIntEvkonyve_38.pdf Download (3MB) | Preview |
Abstract
A solid state amplifier and different measurement set-ups for microelectrode method are investigated by the author. The preamplifier is supplied with a compensating circuit for neutralization of the input shunt capacity, and has unity gain and high input resistance. The different measurement tasks are handled by a bridge which can be connected to the input. In this way, the object under test may be polarized by DC current or stimulated by a square wave. The measurement bridge also serves for measuring the microelectrode resistance. The visual evaluation of the signals is possible by an oscilloscope connected to the output of the low-level preamplifier. The system has also a high-level output for driving signal storage and signal processing devices. The control methods of the measurement set-up parameters are summarized in a Table. The parameters have been controlled by a calibrating circuit which has been also dealt with.
Item Type: | Article |
---|---|
Subjects: | T Technology / alkalmazott, műszaki tudományok > TK Electrical engineering. Electronics Nuclear engineering / elektrotechnika, elektronika, atomtechnika |
Depositing User: | Edina Fejős |
Date Deposited: | 31 May 2024 17:05 |
Last Modified: | 31 May 2024 17:05 |
URI: | https://real.mtak.hu/id/eprint/196250 |