Abstract

The medical application properties of coronary stents describe their behaviour in the human vascular system from planting to functioning. However these properties have great importance to surgeons, not all of them have standardized examination methods. In our study we demonstrate three procedures, which can be used to examine stents, considering the referred properties, like flaring, trackability and MSA (metallic surface area). In the course of our research four stents were investigated, three made of tube and one made of wire, and the results were promising about the application of these methods described in the followings.