Moór, Attila and Müller, T. G. and Kiss, Csaba and Balog, Zoltán and Billot, N. and Marton, Gábor (2014) PACS photometer calibration block analysis. EXPERIMENTAL ASTRONOMY, 37 (2). pp. 225-238. ISSN 0922-6435
|
Text
1310.2730.pdf Available under License Creative Commons Attribution. Download (7MB) | Preview |
Abstract
The absolute stability of the PACS bolometer response over the entire mission lifetime without applying any corrections is about 0.5% (stan- dard deviation) or about 8% peak-to-peak. This fantastic stability allows us to calibrate all scientific measurements by a fixed and time-independent re- sponse file, without using any information from the PACS internal calibration sources. However, the analysis of calibration block observations revealed clear correlations of the internal source signals with the evaporator temperature and a signal drift during the first half hour after the cooler recycling. These effects are small, but can be seen in repeated measurements of standard stars. From our analysis we established corrections for both effects which push the stability of the PACS bolometer response to about 0.2% (stdev) or 2% in the blue, 3% in the green and 5% in the red channel (peak-to-peak). After both corrections we still see a correlation of the signals with PACS FPU tempera- tures, possibly caused by parasitic heat influences via the Kevlar wires which connect the bolometers with the PACS Focal Plane Unit. No aging effect or degradation of the photometric system during the mission lifetime has been found.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Long term behaviour of bolometers; Calibration block observations; PACS bolometers; |
Subjects: | Q Science / természettudomány > QB Astronomy, Astrophysics / csillagászat, asztrofizika |
SWORD Depositor: | MTMT SWORD |
Depositing User: | MTMT SWORD |
Date Deposited: | 26 Mar 2024 15:45 |
Last Modified: | 26 Mar 2024 15:45 |
URI: | https://real.mtak.hu/id/eprint/191043 |
Actions (login required)
Edit Item |