Maróti, Boglárka and Révay, Zsolt and Szentmiklósi, László and Kleszcz, Krzysztof and Párkányi, Dénes and Belgya, Tamás (2018) Benchmarking PGAA, in-beam NAA, reactor-NAA and handheld XRF spectrometry for the element analysis of archeological bronzes. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 317 (2). pp. 1151-1163. ISSN 0236-5731 (print), 1588-2780 (online)
|
Text
JRNC-BCR-691-REVISION_OPEN_ACCESS.pdf Download (943kB) | Preview |
Abstract
Novel detection strategies were developed to improve the PGAA methodology in the analysis of bronze and brass objects. Instrumental neutron activation analysis and X-ray fluorescence were also involved in this study for comparison. The approach was tested on a set of IRMM-certified copper alloys. Reference values for the concentrations of four major additive components (As, Zn, Sn, and Pb) were compared with the results obtained by the involved methods. The bias observed between the reference Sn values and the initial PGAA results was successfully addressed by considering the interference of this line with low intensity copper prompt gamma peaks. In this report results are presented, and the analytical merits are compared.
Item Type: | Article |
---|---|
Subjects: | Q Science / természettudomány > QD Chemistry / kémia > QD01 Analytical chemistry / analitikai kémia |
Depositing User: | Dr. László Szentmiklósi |
Date Deposited: | 27 Sep 2018 07:49 |
Last Modified: | 05 Apr 2023 07:44 |
URI: | http://real.mtak.hu/id/eprint/85501 |
Actions (login required)
![]() |
Edit Item |