Márton, Gergely and Fekete, Zoltán and Fiáth, Richárd and Baracskay, Péter and Ulbert, István and Juhász, György and Battistig, Gábor and Pongrácz, Anita (2013) In vivo measurements with robust silicon-based multielectrode arrays with extreme shaft lengths. IEEE SENSORS JOURNAL, 13 (9). pp. 3263-3269. ISSN 1530-437X

[img]
Preview
Text
IEESens_InVivo_13_3263_proof.pdf

Download (3MB) | Preview

Abstract

In this paper, manufacturing and in vivo testing of extreme-long Si-based neural microelectrode arrays are presented. Probes with different shaft lengths (15–70 mm) are formed by deep reactive ion etching and have been equipped with platinum electrodes of various configurations. In vivo measurements on rats indicate good mechanical stability, robust implantation, and targeting capability. High-quality signals have been recorded from different locations of the cerebrum of the rodents. The accompanied tissue damage is characterized by histology.

Item Type: Article
Uncontrolled Keywords: Neural microelectrode array, deep-brain electrode, silicon-based microelectromechanical system, spike sorting.